Updated: 4/26/2009 7:00:09 AM
(HealthNewsDigest.com) - Just in time for Christmas,—on December 22, 2008—the US Department of Veterans Affairs issued a Patient Safety Alert entitled "Improper set-up and reprocessing of flexible endoscope tubing and accessories." The major finding was that certain tubing modified with an incorrect connector was used to attach an Olympus flexible endoscope to its irrigation source.

This resulted in a backflow of body fluids into irrigation source tubing—where backflow should not occur—during an endoscopic procedure (i.e. a colonoscopy). In other words, supposedly clean irrigation solutions were being contaminated with the patient's body fluids, promoting cross-contamination.

It was also discovered, and reported in the same Patient Safety Alert, that a related component of the endoscopes was being reprocessed (sterilized or disinfected) only at the end of the day rather than after each patient use, as is required by the manufacturer's reprocessing instructions.

Per the Alert, all VA facilities were to investigate their procedures and report back. Media sources indicated that the problems described occurred at the VA hospital in Murfreesboro, TN.

At the heart of the matter are two particular tubes, used in both the actual patient exams and in reprocessing. The tubes are different in appearance: Tube MH-974 is about 10 inches in length (254 mm), while Tube MAJ-855 is around 4 feet (1.2 m) long. Each tube is provided with its own specific connector. MH-974's connector does not contain a valve, while MAJ-855's connector contains an internal one-way valve.

The connectors are very similar in appearance, except for one "wing" in MH-974's, and two wings in MAJ-855's. The connectors are labeled to indicate which tube they belong to.

As Olympus describes it in their "Important Safety Notice" of January 26, 2009: The one-way valve in the proximal connector of the MAJ-855 prevents backflow of fluids from the auxiliary water channel of endoscope into the flushing pump tubing. The one-way valve also prevents water from flowing out of the proximal connector of the MAJ-855 when disconnecting the syringe (used in manual reprocessing) or the flushing pump tubing from the MAJ-855.

On February 13, 2009, the Murfreesboro facility sent letters to nearly 6,400 veterans warning that improperly assembled colonoscopy equipment may have exposed them to Hepatitis B, Hepatitis C, and HIV. This improper assembly could have occurred anytime between April 23, 2003 and Dec. 1, 2008.

From March 8-14, 2009, the VA initiated a "step-up" program whereby all facilities were to check if they have contamination problems, and to arrange new training programs. Based on this new inspection protocol, the Miami VA Medical Center discovered that it, too, has endoscope reprocessing issues. On March 23, Miami VAMC sent letters to about 3,260 veterans, warning that if they had colonoscopies at the facility, improperly sanitized equipment might have exposed them to Hepatitis B, Hepatitis C, or HIV.

By April 17, one case of HIV, five of Hepatitis B, and 11 of Hepatitis C were reported for Murfreesboro. Miami results so far indicate one case of HIV, none of Hep B, and seven of Hep C.

The VA said that there was no way to prove that the patients contracted the illnesses because of treatment at their facilities. However, I would point out that genome analysis (in effect, forensics on the pathogen) and a recent history on the affected individuals would remove most of the doubt. If DNA analysis (RNA in the case of Hep C) showed the same genome, and the patient showed no risk factor other than the colonoscopy, one could be reasonably sure of the source of the infection.

What about the notion that "this improper assembly could have occurred anytime between April 23, 2003 and Dec. 1, 2008"? That's a time span of more than five years. What could have gone wrong?

It is noted that the Olympus endoscope-reprocessing manual recommends the use of the MAJ-855 auxiliary water tube to manually process a portion of the endoscope called the auxiliary water channel. But, medical facilities have on hand expensive units called automated endoscope reprocessors (AERs). Why do something by hand if you have a machine?

Sources tell us that one AER manufacturer informed its customers that MAJ-855 could be used for automatic reprocessing, as long as the connector for MH-974 (with no internal valve) was used with it.

Fair enough, but this is an accident waiting to happen, is it not? For patient use, MAJ-855 must be provided with its proper connector, containing that one-way valve. Otherwise, dangerous backflow is allowed. And, since these two connectors have a similar appearance, one can virtually guarantee that on occasion, the proper connector will not be put back on tube MAJ-855.

Based on this scenario, it is quite unlikely that the problems will be limited to only VA facilities.

Before blame is assessed, and there is surely more than enough to go around, we might ask where the regulatory agencies and non-governmental organizations were, that supposedly are watching over us in these matters.

Michael D. Shaw
Exec VP
Interscan Corporation

mds1@gasdetection.com
http://www.gasdetection.com

www.HealthNewsDigest.com


Read Original Article

© 2009 HealthnewsDigest.com

Customer Care - 1-888-632-5353 Toll Free

  
Additional Articles
Global Warming: Our Best Guess is Wrong...
Wheaties Celebrates 85 Years
Plan B® One-Step—New FDA-Approved ...
Keeping Your Home Flea Free
Web Site Security
Considering A Career In Cosmetology
Knowing How To Handle Meat And Poultry
Pesticide Levels in Blood Linked to ...
Probiotics Help Gastric-bypass Patients...
New Roadside Survey Shows Steady in Use...
FDA Issues Draft Guidance for Industry ...
Ten Tips to Help Your Child Prepare for...
Loneliness: Human Nature and the Need ...
Risks of Delaying ACL Reconstruction in...
Doctors Talk Frankly About What and of ...
Regular Moderate Alcohol Intake Has in ...
Is it Heat or Ice On an Injury?
Do I Need a Supplement?
Put America on a Diet
Train Travel - Less Polluting?
Population Growth and Climate Change
CPR Mattress: An Innovation That Can ...
An Herbicide Used To Protect Corn Also ...
Milwaukee VA Hospital Gains New Spinal ...
Ugandan Medical Male Circumcision to ...
Entertaining Made Simple And Delicious ...
Leading Pioneers in Antibiotic-Free New...
Why Toxicology Testing Takes Time to of...
Segments of Carotid Artery Respond to ...
How to Prevent the Recession from You ...
Arthritis: Can the Body’s Own Defense ...
Stress and Depression Worsen Childhood ...
FDA Takes Enforcement Action Against ...
Parent's Guide to Cord Blood Foundation...
Women Cheat Too (that may include your ...
Proper Fit and Use of Walkers and Canes...
Nine Supplements for Osteoarthritis ...
The Eco-Dentistry Association™ a Week ...
The Metabolism Miracle
University of Phoenix Expands Advanced ...

Add to Google MSN HealthnewsDigest.com
 Add HealthnewsDigest.com
 To My Yahoo  Subscribe with Bloglines   Subscribe in NewsGator Online HealthnewsDigest.com
 News Feed

The materials and articles published on DentalPlans.com are for informational purposes only. Although DentalPlans.com strives to be accurate and complete, the information is provided without liability for errors. DentalPlans.com does not warrant the accuracy or completeness of the information, text graphics, links, or other items contained on DentalPlans.com.

DentalPlans.com expressly disclaims liability for errors or omissions in these materials and DentalPlans.com makes no commitment to update the information on DentalPlans.com.

DentalPlans.com expressly disclaims all liability for the use or interpretation by others of information on DentalPlans.com. Decisions based on information contained on DentalPlans.com are the sole responsibility of the visitors, and visitors agree to hold DentalPlans.com and its Affiliates harmless against any claims for damages arising from decisions visitors make on such information.

Nothing on DentalPlans.com constitutes medical advice or other forms of advice. DentalPlans.com assumes no responsibility for material created or published by third parties linked to DentalPlans.com with or without DentalPlans.coms knowledge.

Terms of Use | Privacy PolicySite Map | Newsletter | Info to Go | DP Goes Green | Affiliate Program | Contact Us |

The DENTALPLANS.COM website is administered by DENTALPLANS.COM, INC., a licensed Florida Discount Medical Plan Organization, 8100 S.W. 10th Street Suite #2000, Plantation, FL 33324. Plans and Programs offered by DentalPlans.com are not health insurance policies. Plans and Programs offered by DentalPlans.com provide discounts at certain health care providers for medical services. Plans and Programs offered by DentalPlans.com do not make payments directly to the providers of medical services. The Plan or Program member is obligated to pay for all health care services but will receive a discount from those health care providers who have contracted with the Plan, Program or discount plan organization.

© 1999-2010 DentalPlans.com, Inc. All Rights Reserved. Patents Pending.

BBBOnLine Reliability Seal    HACKER SAFE certified sites prevent over 99% of hacker crime.